LogTag® UTRIX-16 is a multi-trip temperature logger with integrated USB port and automatic PDF reporting capabilities, designed for temperature monitoring in high-precision cold supply chains such as pharmaceuticals, vaccines, biological products, and food.
| Product Model | UTRIX-16. |
| Sensor Measurement Range | -30°C to +70°C (-22°F to +158°F). |
| Operating Temperature Range | -30°C to +70°C (-22°F to +158°F).* |
| Storage Temperature Range | 0°C to +40°C (32°F to +104°F). |
| Rated Temperature Reading Accuracy | Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F). Better than ±0.7°C (±1.3°F) for -30°C to -20°C (-22°F to -4°F) & +40°C to +60°C (+104°F to +140°F). Better than ±0.8°C (±1.5°F) for +60°C to +70°C (+140°F to +158°F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration. |
| Rated Temperature Reading Resolution | Less than 0.1°C/F. LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®. |
| Sensor Reaction Time | Typically less than 7 minutes (T90) in moving air (1m/s) in EN12830:1999. |
| Recording Capacity | 16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging. |
| Sampling Interval | Configurable from 30 seconds to 18 hours. |
| Logging Start Options | Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time. |
| Recording Indication | Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator. |
| Alarms | 1 configurable upper and 1 configurable lower alarms. |
| Download Time | Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report. Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured). |
| Environmental | IEC 60529: IP64 with USB cap fitted. |
| Power Source | CR2450 3V LiMnO2 Battery (Fixed). |
| Battery Life | 1 year storage, followed by 2 – 3 years of normal use (based on 15 minute logging, download data monthly). |
| Real Time Clock | Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C). |
| Connection Interface | USB 2.0, A-type plug |
| Software | PDF Reader, LogTag® Analyzer 2.5 or higher. |
| Multi-language Support | Multi-language support for English, Spanish, French, Russian and German. |
| Size | 93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap. |
| Weight | 36g. |
| Case Material | Polycarbonate. |
| Footnote | *(-30°C to +70 °C applies to Loggers with serial numbers 5x40x44197 or greater. Loggers with serial numbers prior to this have an operating temperature range of -25°C to +70°C). |
UTRIX-16 User Manual | |
English Version | VietNamese Version |